Update to the newest device firmware
Summary
Perform a firmware update on the CURE.LOG using a paperclip and a USB-C cable.
Cause
- Bug fixes
- Performance improvements
- New features
Preconditions
Before proceeding, ensure you have the following:
- Check the installed firmware version
(refer to the "info.txt" file on the CURE.LOG device) - Paperclip (for button press)
- USB-C cable (for connection)
- Downloaded firmware file in .uf2 format
Remedy
- Plug one end of the USB-C cable into the computer.
- Unplug the CURE.LOG device.
- Connect CURE.LOG device via USB-C.
Locate the small reset button on the CURE.LOG device.
Use a paperclip to gently press and hold the button.
Plug the other end of the USB-C cable into the CURE.LOG device.
Release the button after a 2 seconds. - Access Device in File Explorer:
Open your file explorer (e.g., Windows Explorer).
Look for the device labeled as "RPI-RP2."
If "RPI-RP2" does not appear, restart with step 2
(Button was not pushed properly or device was still powered on) - Drag and Drop Firmware File:
Locate the downloaded firmware file (.uf2) on your computer.
Drag and drop the file onto the device drive (RPI-RP2). - Initiate Update:
The firmware update process will begin automatically.
Wait for the update to complete.
The device will restart upon completion. - Ready to Use:
After the restart, the device will reappear as "CURELOG" in your file explorer.
Your CURE.LOG device is now updated and ready for use.