Update to the newest device firmware


Summary

Perform a firmware update on the CURE.LOG using a paperclip and a USB-C cable.

Cause

  • Bug fixes
  • Performance improvements
  • New features

Preconditions

Before proceeding, ensure you have the following:

  • Check the installed firmware version
    (refer to the "info.txt" file on the CURE.LOG device)
  • Paperclip (for button press)
  • USB-C cable (for connection)
  • Downloaded firmware file in .uf2 format

Remedy

  1. Plug one end of the USB-C cable into the computer.
     
  2. Unplug the CURE.LOG device.
     
  3. Connect CURE.LOG device via USB-C.

    Locate the small reset button on the CURE.LOG device.
    Use a paperclip to gently press and hold the button.
    Plug the other end of the USB-C cable into the CURE.LOG device.
    Release the button after a 2 seconds.

  4. Access Device in File Explorer:

    Open your file explorer (e.g., Windows Explorer).
    Look for the device labeled as "RPI-RP2."
    If "RPI-RP2" does not appear, restart with step 2
    (Button was not pushed properly or device was still powered on)

  5. Drag and Drop Firmware File:

    Locate the downloaded firmware file (.uf2) on your computer.
    Drag and drop the file onto the device drive (RPI-RP2).

  6. Initiate Update:

    The firmware update process will begin automatically.
    Wait for the update to complete.
    The device will restart upon completion.

  7. Ready to Use:

    After the restart, the device will reappear as "CURELOG" in your file explorer.
    Your CURE.LOG device is now updated and ready for use.