The PowerCURE™ is running and heating, but the display shows no current or power
Cause:
There might be a loose connection!

Remedy:
- Check the connection of the plug J1 on the main board (see picture).
 - Check the voltage on this plug while running:
- white -> black = 0.5 – 1.0 VDC
 - green – > black = 15 VDC
 - brown -> black = -15 VDC
 
 
- If you don’t have 0.5-1.0V, please check the wiring of this cable, which goes to a small board inside the of the Pacemaker HF.
 
NOTE:
Check this plug only when the unit is switched off and discharged!

		