Phase voltage error or the voltage value on the touch screen is much lower than on the main terminal


 

Summary

If the voltage value drops below 350 V the error message “Phase voltage error” shows up.

  • Check the main supply: L1 / L2 / L3.

 

  • Check also the input directly inside the PM.

 

Defect of the QM board


The voltage error could also be a defect in the QM board.

  • The QM board receives it’s supply from the electric board of the PM.
  • Remove the supply cable to check.
  • The voltage value should go up around 20V (e.g. 375V to 395V).
    • If the increase is bigger than 20 V, the QM board has a defect and needs to be replaced.
  • The electrical balancing of the two values has been set by CANMAN. A discrepancy can be adjusted.

 

  • lose the small electronic board of Qualimaker (if applicable).
    • This board is supplied by the big electronic board of PM.
  • Remove the supply cable.
    •  the voltage value goes up around 20 V (e.g. 365V to 385V).
  • Measure the input of terminals, for example 385V.
  • Turn the small blue potentiometer (picture below).
  • Control the value on touch screen.
    • Must be around 20V more than the measured value, if you have a QM board.
  • When adjusted, connect the supply of QM.
    • The correct value should occur on touch screen.

NOTE

The V4 mainboard has two potentiometers (see picture below).

  • Adjust the “UMains” potentiometer in this case.

Trouble Shooting

For the correct function of Qualimaker check the 3 LED’s (see picture below):

  • LD16 = state of profibus connection, must be on at all time.
  • LD1 = program state, must be on at all time.
  • LD2 = clock(cycle) of program, must be blinking.


 